Bio-Rad Trans-Blot® Plus Cell User Manual
Page 16

Table 3.3 Isoelectric Focusing, Native, Acid Urea Gels, Basic
Proteins
These conditions were determined empirically using 12.5% Tris-HCl Criterion
gels and native horse myoglobin samples. Read about acidic transfers in Section 2.7
of this manual.
High Intensity Field Conditions (Cooling Required)
PLATE ELECTRODE DISTANCE
4 cm
7 cm
10 cm
Buffer Power
Run
Power
Run
Power
Run
conditions
Time conditions Time conditions Time
0.7% Acetic
50V/2A
45 min
70V/2A
60 min
100V/1.9A
60 min
acid , pH 2.8
70V/3A
15–30 min
110V/3A
15–30 min
125V/2.4A 15–30 min
Standard Field Conditions (Cooling Recommended)
PLATE ELECTRODE DISTANCE
4 cm
7 cm
10 cm
Buffer Power
Run
Power
Run
Power
Run
conditions Time conditions Time conditions Time
0.7% Acetic
10V/0.4A
16 Hrs.
10V/0.25A
16 Hrs.
10V/0.15A
16 Hrs.
acid , pH 2.8
20V/0.8A
20V/0.55A
20V/0.35A
30V/0.8A
30V/0.55A
14