Bio-Rad ReadyPrep™ 2-D Starter Kit User Manual

Page 25

Advertising
background image

23

4. - The first line displays the step number, slope and voltage

- Select HRS:MIN (default)
- Enter time of 20 min.
- Press NEXT

S 01

250 V

HRS:MIN

vhours

>

ENTER TIME:

00:20

>

NEXT

>

5. - Select ADD (default)

- Enter Step 2
- Press NEXT

ADD inset delete >
ENTER STEP #

2 >

NEXT

>

METHOD COMPLETE

>

6. - Enter step 2 final voltage

7 cm: 4,000 V; 11 cm: 8,000 V; 17 cm: 10,000 V)

- Select LINEAR ramp. Use key next to arrow, >, to move

the brackets, [ ] ,around the selection

- Press NEXT

S 02

xxxx V

>

SELECT VOLTAGE SLOPE
LINEAR

[

д

д]

>

NEXT

>

7. - The first line displays the step number, slope and voltage

- Select HRS:MIN (default)
- Enter time

(7 cm: 2:00; 11 cm: 2:30; 17 cm 2:30)

S 02

xxxx V

>

HRS:MIN

vhours

>

ENTER TIME:

xx:xx >

NEXT

>

8. - Select ADD (default)

- Enter step 3
- Press NEXT

ADD Inset delete >
ENTER STEP #

3 >

NEXT

>

METHOD COMPLETE

>

9. - Enter step 3 final voltage

(7 cm: 4,000 V; 11 cm: 8,000 V; 17 cm: 10,000 V)

- Select RAPID ramp. Use key next to arrow, >, to move

the brackets, [ ] ,around the selection

- Press NEXT

S 03

xxxx V

>

SELECT VOLTAGE SLOPE
RAPID

[

]

>

NEXT

>

- The first line displays the step number, slope and

voltage

- Select VHOURS (default)
- Enter time (volt hours)

(7 cm: 10,000; 11 cm: 20,000; 17 cm 40,000)

- Press NEXT

S 03

xxxx V

>

HRS:MIN

VHOURS

ENTER TIME:

xxxxx

>

NEXT

>

10.

Advertising
This manual is related to the following products: