Bio-rad protocols – Bio-Rad PROTEAN® i12™ IEF System User Manual
Page 50

Appendix C IEF Protocols
46 |
PROTEAN i12 IEF System Instruction Manual
46 |
Bio-Rad Protocols
(All protocols have a 50 μA current limit, though 100 μA is possible.)
Step
Voltage (V)
Ramp
Time
Units
1
4,000
Rapid
15,000
Volt Hr
2
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:15
HH:MM
2
4,000
Gradual
1:00
HH:MM
3
4,000
Rapid
15,000
Volt Hr
4
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
4,000
Rapid
10,000
Volt Hr
2
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:15
HH:MM
2
4,000
Gradual
1:00
HH:MM
3
4,000
Rapid
20,000
Volt Hr
4
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:15
HH:MM
2
4,000
Gradual
1:00
HH:MM
3
4,000
Rapid
25,000
Volt Hr
4
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
4,000
Rapid
16,000
Volt Hr
2
500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:15
HH:MM
2
4,000
Gradual
1:00
HH:MM
3
4,000
Rapid
16,000
Volt Hr
4
500
Hold
7cm pH5.5–6.7
7cm pH6.3–8.3
7cm pH7–10 R
7cm pH7–10 G
7cm pH3–10 R
7cm pH3–10 NL R
7cm pH4–7 R
7cm pH5–8 R
7cm pH3–10 G
7cm pH3–10 NL G
7cm pH4–7 G
7cm pH5–8 G
7cm pH3–6 R
7cm pH3.9–5.1
7cm pH4.7–5.9
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:15
HH:MM
2
4,000
Gradual
1:00
HH:MM
3
4,000
Rapid
20,000
Volt Hr
4
500
Hold
7cm pH3–6 G