Bio-Rad PROTEAN® i12™ IEF System User Manual
Page 53

Appendix C IEF Protocols
PROTEAN i12 IEF System Instruction Manual
| 49
Step
Voltage (V)
Ramp
Time
Units
1
10,000
Rapid
60,000
Volt Hr
2
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:30
HH:MM
2
10,000
Gradual
2:00
HH:MM
3
10,000
Rapid
60,000
Volt Hr
4
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
10,000
Rapid
44,000
Volt Hr
2
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:30
HH:MM
2
10,000
Gradual
2:00
HH:MM
3
10,000
Rapid
70,000
Volt Hr
4
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:30
HH:MM
2
10,000
Gradual
2:00
HH:MM
3
10,000
Rapid
88,000
Volt Hr
4
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
10,000
Rapid
63,000
Volt Hr
2
1,500
Hold
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:30
HH:MM
2
10,000
Gradual
2:00
HH:MM
3
10,000
Rapid
63,000
Volt Hr
4
1,500
Hold
24cm pH5.5–6.7
24cm pH6.3–8.3
24cm pH7–10 R
24cm pH7–10 G
24cm pH3–10 R
24cm pH3–10 NL R
24cm pH4–7 R
24cm pH5–8 R
24cm pH3–10 G
24cm pH3–10 NL G
24cm pH4–7 G
24cm pH5–8 G
24cm pH3–6 R
24cm pH3.9–5.1
24cm pH4.7–5.9
Step
Voltage (V)
Ramp
Time
Units
1
250
Rapid
0:30
HH:MM
2
10,000
Gradual
2:00
HH:MM
3
10,000
Rapid
44,000
Volt Hr
4
1,500
Hold
24cm pH3–6 G