Bio-Rad PowerPac™ Universal Power Supply User Manual
Page 11
Step Procedure
Description
4.
"Edit HV Method" screen
Enter name using the alphanumeric PDA
pad
Rename—check "rename" to overwrite
an existing method with the new name.
Save—saves the edited methods in the
designated methods list
Delete—deletes the selected method.
Beam—beams the selected method to
the targeted PowerPac power supply
5.
New Method steps
New Method:
Select "S1---Blank Step---" from the
Method Steps drop down menu
Existing method steps:
Existing / saved method:
Select a step to edit from the Method
Steps drop-down menu
Click the down arrow in the lower right
hand corner of the step list to access
additional steps
6.
PowerPac HV
PowerPac HV : Basic Method
Basic Method
Const Mode—select voltage, current, or
power as the constant parameter
Parameter Values—enter value for
constant parameter and values for limit
parameters.
Timing—select time (HHMM), volt-hours,
or untimed step option
Rename—you can rename the saved
method, or duplicate the method by
saving it under a new name (uncheck
rename)
Save—select to save changes
8