Bio-Rad PowerPac™ Universal Power Supply User Manual

Page 12

Advertising
background image

Step Procedure

Description

6. (

Continued)

Delete—select to delete method
Beam—transmits saved method to PowerPac
power supply
Delete Step—deletes the displayed step (no
warning message)
Methods—returns to method screen

7.

PowerPac HV

PowerPac HV: IEF Method

IEF Method

IEF Methods are set to run at constant voltage
with current and power as the limiting
parameters. The maximum voltage value
programmed can be reached via three different
ramping algorithms: linear, rapid, or slow.

Ramp—select voltage ramping: L (linear),
R (rapid), or S, (slow)
Parameter Values—enter values for
constant voltage, current limit, and power
limit
Timing—select time T (HHMM) or Vh
(Volthours)
Delay—check delay to program a timed-
delayed first step

Delay option for Step 1

Step 1 – Delay option.
This option is available as the first step only
for PowerPac HV IEF methods.

Check delay.
Enter time for the delay step (HHMM).

8.

PowerPac Univeral

PowerPac Univeral Method

Method

Const Mode—select voltage, current, or
power as the constant parameter
Parameter Values—enter value for
constant parameter, and values for limit
parameters
Timing—select time (HHMM), Volt-hours
Rename—check Rename to rename the
saved method, or uncheck Rename to
duplicate the method by saving it under a
new name
Save—select to save changes

Select another step to edit from the Step
drop-down list and edit the step.

9

Method: SDS-PAGE

Step # 1

Const Mode

Rename

Volts: 100

mAmps: 2500

Watts: 500

Vhours: 00000

Untimed

Save

Delete

Beam

Methods

Delete Step

V

A

Timing

T

Vh

W