Troubleshooting guide – Bio-Rad Human Metabolic and Hormone Assays User Manual

Page 26

Advertising
background image

24

Possible Causes

High Inter-Assay CV


Standards were not reconstituted
consistently between assays

Reconstituted standards and
diluted samples were not stored
properly

Bottom of filter plate not dry

Possible Solutions

Incubate the reconstituted standards
for for the recommended time
period. Always be consistent with the
incubation time and temperature.

Diluted samples should be prepared
on ice as instructed. Prior to
plating, the reconstituted standards
and diluted samples should be
equilibrated to room temperature.

Dry the bottom of the filter plate with
absorbent paper towel (preferably
lint-free) to prevent cross-well
contamination.

Troubleshooting Guide

This troubleshooting guide addresses problems that may be encountered
with Bio-Plex Pro

RBM assays. If you experience any of the problems

listed below, review the possible causes and solutions provided. Poor
assay performance may also be due to the Bio-Plex

®

suspension array

reader. To eliminate this possibility, use the validation kit to validate all the
key functions of the array reader and to assist in determining whether or
not the array reader is functioning properly.