2. Equipment Set-up and Installation

2.1 Unpacking the xMark

Prior to opening the box, check for any outer damage. Document any damage, if present.

Place the box upright with the TOP label facing up and open the box. Remove the instrument from
the box in this position only.

Warning! Grasp the instrument firmly under the right and left sides before attempting to lift or carry
the instrument out of the box.

Check the instrument for damaged or broken parts. Document any damage in detail.

Each xMark instrument includes the following accessories. Verify that they are present:

• 1 Microplate Manager (MPM) software with manual (packaged separately)

• 1 USB cable

• 1 power cord

• 1 Dust cover
• 2 Spare fuses

• 1 User manual
• 1 Warranty card

Verify that the serial number on the back of the instrument is the same as that on the outside of the
box.

Located on the underside of the instrument is a docking bolt that must be unscrewed counter-
clockwise prior to use. When the docking bolt cannot be turned any further, the scan arm is free
inside the instrument.

The USB port is located on the back of the instrument. There is one connector, which allows the user
to connect to one USB instrument.

Before switching the power on for the first time, allow the xMark to warm to room temperature for 6
hours
after unpacking, to prevent any condensation problems.

Plug the included power cord into the main power connector port. Ensure that the instrument is
connected to a main power outlet with a grounded conductor.

2.2

Location of the Instrument

The xMark Microplate Spectrophotometer is a precise optical measuring instrument. The following
conditions must be met to ensure maximum sensitivity:

• Place the instrument on a flat surface that is free of vibration

• Do not place the instrument in direct sunlight

• Make sure that the instrument is operated in a dust-free environment