Bio-Rad Profinia™ Protein Purification Instrument User Manual

Page 2

Advertising
background image

Life Science
Group

06-0784 0107 Sig 1106

Bulletin 5539 US/EG Rev A

Bio-Rad
Laboratories, Inc.

Web site www.bio-rad.com USA 800 4BIORAD Australia 61 02 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 21 3237 9400
Canada 905 712 2771 China 86 21 6426 0808 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00 France 01 47 95 69 65
Germany 089 318 84 0 Greece 30 210 777 4396 Hong Kong 852 2789 3300 Hungary 36 1 455 8800 India 91 124 4029300 Israel 03 963 6050
Italy 39 02 216091 Japan 03 5811 6270 Korea 82 2 3473 4460 Mexico 52 555 488 7670 The Netherlands 0318 540666 New Zealand 0508 805 500
Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 27 861 246 723
Spain
34 91 590 5200 Sweden 08 555 12700 Switzerland 061 717 95 55 Taiwan 886 2 2578 7189 United Kingdom 020 8328 2000

Fig. 2. UV absorbance and conductivity profiles of desalting runs on the
Profinia system.
Absorbance at 280 nm and conductivity readings of BSA
samples were collected over time. (

), UV absorbance at 280 nm; (

),

conductivity trace. The two vertical lines mark the collected protein peak (1D),
which is resolved completely from the salt peak.

screen. To desalt the 2 ml sample, a 2 ml sample loop and
a 10 ml desalting cartridge were used. To desalt the 10 ml
sample, a 10 ml sample loop and a 50 ml desalting cartridge
were used. Profinia software was used for real-time monitoring
of the UV absorbance and conductivity of the sample during
the desalting run and for collection and analysis of run data.
Buffers and desalting cartridges are part of the Profinia
desalting kit, while sample loops can be obtained as an
accessory for the desalting-only application.

Desalting by Dialysis

The dialysis protocol described by Zumstein (1994) was
followed using a Spectra/Por dialysis membrane with a
molecular weight cutoff of 14 kD (Spectrum Laboratories) and
a 1:500 volume of phosphate buffered saline (PBS). During
desalting, PBS was mixed continuously using a magnetic
stirrer, and conductivity of the BSA sample was monitored
at 15 min intervals using a MeterLab CDM210 conductivity
meter (Radiometer Analytical). Dialysis was complete when
the conductivity of the sample reached equilibrium with that
of the exchange buffer.

Results and Discussion

Using the Profinia system, UV absorbance and conductivity
data were automatically collected and plotted to demonstrate
the performance of this desalting method (Figure 2). The
Profinia system also automatically delivered the desalted
protein product to a collection tube in 15 min, regardless
of the sample size being desalted. In contrast, the dialysis
method required manual conductivity measurements to
determine when buffer exchange was complete.

Both methods were capable of desalting protein samples
to completion; however, there were notable differences in
performance (Table 1). The time required to complete desalting
with the Profinia system was 8- to 10-fold less than the time
required for dialysis, depending on the size of the sample.
Although both methods resulted in >90% yield, the final
concentration of the dialyzed protein sample was 2-fold higher.
Finally, the Profinia system used 20-fold less desalting buffer
than was required for dialysis.

Table 1. Performance comparison of the two desalting methods.
Average values from three replicate desalting runs (n = 3) are shown.

Final Protein

Total Time

Concentration

Recovery

Buffer

Method

Required

(mg/ml)

(mg)

Yield

Used

2 ml Sample
Profinia system

15 min

1.4

5.6

93%

50 ml

Dialysis

120 min

2.8

5.6

93%

1,000 ml

10 ml Sample
Profinia system

15 min

1.4

28.0

93%

250 ml

Dialysis

150 min

2.7

28.3

94%

5,000 ml

Conclusions

Dialysis and gel filtration chromatography are two commonly
used methods for desalting protein solutions. As the
mechanism of buffer exchange is not the same, there are
advantages to each method. Dialysis will usually produce a
more concentrated product than gel filtration, though some
sample dilution should still be expected. Desalting by gel
filtration using the Profinia protein purification system offers
substantial time savings as well as the significant advantages
of greatly decreased buffer consumption and an automated,
scalable process.

Reference

Zumstein L, Dialysis and ultrafiltration, appendix 3C.1 in Current Protocols
in Molecular Biology (Ausubel FM et al., eds), Wiley, New York (1994)

Information in this tech note was current as of the date of writing (2006) and
not necessarily the date this version (rev A, 2007) was published.

AU (UV channel 2)

Run time, min

0.4

0.3

0.2

0.1

0.0

12

13

14

15

16

17

18

2 ml BSA

Conductivity

, mS/cm

100

80

60

40

20

AU (UV channel 2)

Run time, min

0.4

0.3

0.2

0.1

0.0

12

13

14

15

16

17

18

10 ml BSA

Conductivity

, mS/cm

100

80

60

40

20

BSA

Salt

BSA

Salt

1D

1D

Advertising