Fig. 2. Separation of hemoglobin from ampholytes
and sucrose on a mixed bed ion exchange column
(AG 501-X8 resin).

A hemoglobin blank in the sucrose

determination gives an absorbance of about 0.1. Neither
sucrose nor ampholyte emerged from the column in the
total volume tested.

3

AG 501-X8 resin has also been shown to be useful

for separating peptides of greater than 4,000 daltons from
ampholytes.

11

Table 3. Deionization of Urea by Batch and
Column Methods

Batch

Column

Sample

100 ml 6 M urea

100 ml 6 M urea

Starting conductivity

70 µmho/cm

70 µmho/cm

Amount of mixed bed resin

5 grams

5 grams (~8 ml)

Final conductivity

5.0 µmho/cm

0.2 µmho/cm

Time

~5 hours

~10 minutes

Ampholyte Removal

Carrier ampholytes may be quantitatively removed

from protein fractions derived from isoelectric focusing
using mixed bed resin. Mixed bed ion exchange chro-
matography represents a method for the quantitative
removal of carrier ampholytes. Figure 2 illustrates the
separation of proteins from ampholytes and sucrose. A 3
ml sample was applied to a 0.9 x 25 cm column of AG
501-X8 resin. AG 501-X8 resin has also been shown to be
useful for separating peptides of greater than 4,000 dal-
tons from ampholytes.

10