Technical Description

Table 1. Bio-Gel P Gel Product Description

Matrix

Bio-Gel polyacrylamide gel

Particle size

Medium

90-180 µm

Fine

45-90 µm

Extra fine

< 45 µm

Shipping medium

Shipped dry

Resistance

pH

2-10

Pressure

15 psi

Organic solvents

< 20%

Working temperature range

4-80 °C

Temperature limits

Autoclavable, at pH 5.5-6.5, at
120 °C for 30 min

Storage

Dry, at room temperature; in dis-
tilled water or aqueous buffers
at 4 °C with 0.02% sodium azide

3

exclusion properties of the gel, and will in some cases enhance
separation of complex mixtures of poorly water soluble small
molecules such as nucleotides, peptides, and tannins. Formamide
may be used at full strength, because Bio-Gel P gel is completely
swelled by this solvent.

Bio-Gel P gel is autoclavable at pH 5.5-6.0 in buffers such as

50 mM HEPES, MES, or citrate at 120 °C for 15-30 minutes. At
room temperature, the recommended operating pH range is 2-10.
Bio-Gel P gel is susceptible to hydrolysis of amide groups at higher
or lower pH. Flow rate and resolution increase with temperature over
the range of 4-80 °C.