Protocol 1
The following method can be used to apply a sample directly to
the ProteinChip SEND ID array for peptide analysis.
1. Mix your sample 1:1 (v/v) with 50% ACN (v/v) and 0.2% TFA

in deionized (DI) water.

2. Add 1–2 µl of sample to each spot on the ProteinChip

SEND ID array.

3. Allow the spots to air-dry.
4. Read the arrays in the ProteinChip SELDI reader.

Protocol 2
The following method can be used to clean up a sample on
the ProteinChip array to remove salt and urea contamination
(see Figure 4). The C-18 chemistry on the surface selectively retains
peptides through their hydrophobic regions while impurities are
washed away.

1. Mix sample with 0.2% TFA at a 1:1 (v/v) ratio.
2. Apply 5 µl of 0.1% TFA to the ProteinChip SEND ID array.

Remove within 30 sec; do not agitate. Repeat.

3. Spot 2 µl of sample onto the array.
4. Incubate the samples for 10 min in a humidity chamber

at room temperature. Remove the samples.

5. Add 5 µl of 0.1% TFA to each spot. Remove within

30 sec; do not agitate.

6. Add 2 µl of 25% ACN (v/v) and 0.1% TFA in DI water to

each spot.

7. Allow the array to air-dry.
8. Read the array in the ProteinChip SELDI reader.