background image

Table 1. Preparation of ProteinChip EAM in solution using
commonly used solvent systems.

EAM

Solvent System

Preparation

ProteinChip CHCA

50% acetonitrile

49.75% HPLC
grade H

2

O

0.25% trifluoroacetic
acid (TFA)

ProteinChip SPA

50% acetonitrile

49.50% HPLC
grade H

2

O

0.50% TFA

ProteinChip EAM-1 50% acetonitrile

49.50% HPLC
grade H

2

O

0.50% TFA

© 2006 Bio-Rad Laboratories, Inc.

Preparation

1. Add 200 µl solvent to vial

(there should be undissolved
CHCA in the bottom of the
tube)

2. Mix

3. Centrifuge

4. Draw off supernatant and

dilute with equal part solvent

1. Add 400 µl solvent to vial

2. Mix

3. Centrifuge

1. Add 200 µl solvent to vial

(there should be undissolved
EAM-1 in the bottom of the
tube)

2. Mix

3. Centrifuge

4. Draw off supernatant and

dilute with 2x volume of
solvent