Bio-Rad Foresight™ Chromatography Columns, Prepacked User Manual

Page 19

Advertising
background image

Technical Support: 1-800-424-6723 • www.bio-rad.com 15

Section 5
Troubleshooting Guide

Fault

Possible cause

Action

Foresight

RoboColumn
units are
clogged

The sample is too
viscous

There is too much
cell debris in the
sample

The column has
been regenerated
and used for too
many cycles

Increase dilution of
the cell isolate before
lysis, or dilute after
the lysation

Centrifuge and/or
filter the sample if
unclarified sample
has been used

Reduce the number
of cycles

Foresight
RoboColumn
units are leaking

Columns are
clogged

The O-ring seal is
damaged

The insertion depth
of the robotic needle
is incorrect

See above

Check the taught
X and Y coordinates
for RoboColumn
labware for precision

Check the insertion
depth (16 mm
from the top of
the column) of the
robotic needle

Foresight
RoboColumn
units are lifted
up from the
96-well array
plate during the
pull-out of the
robotic needles

Columns are not securely
seated into the 96-well
array plate

Make sure that all
columns are securely
seated into the
96-well array plate by
firmly pressing down
before starting the
experiment

Low protein
recovery in the
elution fraction(s)

The residence time for
sample loading is too low

Decrease the flow rate or
increase residence time
for sample loading