Cooper Instruments & Systems LTL 730 Wireless Tension Link Load Cell User Manual

Page 22

Advertising
background image

CF 150 LTL 730

19

12/08

Units

Press this key to set the following:
Choice of power-up unit of measure
Enable or disable lbf, kgf, N, custom unit 1 and 2

Power

Press this to set the following:
Enable Auto-shutdown (Y/N)
Shutdown Timer (set minutes before idleness causes unit to shut off)

Chpwd

Press this to change the Config menu password.

Reset

Press and you are given the choice of resetting the Communicator to its factory defaults.

9.4 Communicator Print Formats

Press the Ptfmt soft key to select the print format that will be sent from the Communicator on Com 1 to a
peripheral printer when you press the Print soft key. Choose from the formats below. Data sent is shown below.
The default is format #1.

Format #1

Current displayed value (peak or live force) plus unit of measure on the active LTL 730 only.
104.5 lbf (live force example)
302.5 lbf (peak force example)

Format #2

Live force value plus unit of measure on 1st line and Peak force and unit of measure on 2nd line
for active LTL 730 only.
104.5 lbf
302.5 lbf

Format #3

Live force value plus unit of measure on 1st line and Peak force and unit of measure on 2nd line
with descriptive prefixes on each line for active LTL 730 only.
Force 104.5 lbf
Peak 302.5 lbf

Format #4

Fixed position output for active LTL 730 only. Works well with RS- 232 utilities, such as
WedgeLink, for parsing into a spreadsheet such as Microsoft Excel.

Excel

A B

1

104.5 lbf

2

302.5 lbf

Fixed

characters.

Position Description
1-8

Live force number

10-16

Displayed unit of measure (up to 7 characters)

18-25

Peak force number

27-33

Displayed unit of measure (up to 7 characters)

9,17,26 Commas
34 Carriage

Return

Example 1

<sp><sp><sp>104.5,<sp><sp><sp><sp>lbf,<sp><sp><sp>302.5,<sp><sp><sp><sp>lbf<CR>


Example 2 (custom unit)

<sp>140000.,<sp><sp><sp><sp><sp>kg,<sp>165450., ,<sp><sp><sp><sp><sp>kg<CR>


This would yield (shown in courier):

111111111122222222223333333333444444444455555555

123456789012345678901234567890123456789012345678901234567

104.5, lbf, 302.5,

lbf<CR>

140000., kg,

165450.,

kg<CR>

Format #5 Live force, unit of measure, peak force, unit of measure for active LTL 730 only. All tab separated.

104.5<tab>lbf<CR> (if presently displaying live readings)
302.5<tab>lbf<CR> (if presently displaying peak readings)

Format #6 Cell #, description, live force, unit of measure for all active LTL 730s plus a total.