7 – Troubleshooting

E5 – File Operations

Error Number

Description

Corrective Action

502

File operations communication
error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2
5. If the error persists contact QBC Technical Support

503 thru 506

Software configuration file read or
write error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

518

No memory available error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

E6 – Runtime Functions/Routines

Error Number

Description

Corrective Action

600

General runtime error

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Turn the instrument Power (Ref. page 4-2)
5. If the error persists contact QBC Technical Support

601

Internal temperature of the
instrument has exceeded 45 ºC
(113 ºF)

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Tip the instrument up and check for obstructions in the

exhaust port and clear is present

5. Confirm the instrument is set-up properly (Ref. page 2-1)

a. Ensure there is at least 2 inches of clearance around the

instrument for airflow

b. Ensure adequate area above the device for airflow
c. Ensure the instrument is not being operated outside of the

environmental specifications (Ref. page 2-3)

6. Power down the instrument (Ref. page 4-10)
7. Turn the instrument Power on (Ref. page 4-2)
8. if the error persists contact QBC Technical Support

602

Internal temperature of the
instrument has fallen below 15 ºC
(59 ºF)

1. Record the error message displayed on the LCD screen
2. Export the support file (Ref. page 2-8)
3. Power down the instrument (Ref. page 4-10)
4. Tip the instrument up and check for obstructions in the

exhaust port and clear is present

5. Confirm the instrument is set-up properly (Ref. page 2-1)

a. Ensure the instrument is not being operated outside of the

environmental specifications (Ref. page 2-3)

6. Power down the instrument (Ref. page 4-10)
7. Turn the instrument Power on (Ref. page 4-2)
8. if the error persists contact QBC Technical Support