Drucker Diagnostics QBC STAR Dry Hematology Analyzer User Manual

Page 68

Advertising
background image

QBC Star System Operator’s Manual

6000-300-000 7–18

16

Could not find Control / PLT interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close the lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support

17

Could not find PLT / Plasma interface

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

4. Close the lid and rescan the sample
5. If the problem persists, draw a new sample and retest
6. If the problem persists after retest, create a new support file

and contact Technical Support

18

Could not find float top

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid and visually check sample

a. If the float did not descend or the float top is not observed,

draw a new sample and retest

b. If the float is descended or the float top is observed, rotate

the sample 180 degrees and rescan the sample

4. If problem persists, create a new support file and contact

Technical Support

19

Could not find meniscus

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid and visually check sample

a. If no visible meniscus is observed, draw a new sample and

retest

b. If a visible meniscus is observed, rotate the sample 180

degrees and rescan the sample

4. If the problem persists, create a new support file and contact

Technical Support

20

Bad fill line length

1. Record the error message on the LCD display
2. Create a support file (Ref. page 2-8)
3. Open the lid and visually check sample

a. If the fill lines are not visible or are scratched off, draw a

new sample and retest

b. If the fill lines are visible, rotate the sample 180 degrees

and rescan the sample

4. If the problem persists, create a new support file and contact

Technical Support

22

Unreadable Grans layer (L3)

1. Record the error message on the LCD display
2. Record the age of the blood sample
3. Create a support file (Ref. page 2-8)
4. Open the lid, rotate the sample 180 degrees and reinsert it

into rotor

5. Close the lid and rescan the sample
6. If the problem persists, draw a new sample and retest
7. If the problem persists after retest, create a new support file

and contact Technical Support