5 - empty dispenser – Hudson Robotics Micro10x User Manual
Page 13
Advertising

13
Figure 2. 4.1
Volume – This is the total volume that will be primed through the system. Enter a
value between 120 and 9999uL.
Speed – This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation. Do not exceed more
than 30% speed on any value over 1000ul.
Tip Touch – Activate or deactivate tip touching after each dispense.
Tip Touch – Adjust where in the Priming trough you would like to touch the tips
after a prime. This is helpful when running viscous liquids.
2.5 - Empty Dispenser
This feature reverses the pump, removing most of the liquid from the manifold
and tubing and returns it to the source container (Figure. 2.5 1).