3.2 - Disassembly of the Pump Head Module


After the pump head module is removed from the pump you will able to access
the Teflon adapters. The adapters lock the liner and piston into the proper
position inside the cylinder case. You will remove the adapters by using a 5/8”
open ended wrench and turning the adapters counter clockwise until they are
removed from the cylinder case (Figure 3.2.1). When the adapters are removed
you will notice that there are two flat surfaces machined in the liner that need to
be re-aligned at the correct location before the adapters are re-installed (Figure
3.2.2). Now remove the pump bracket from the pump cylinder by removing the
flathead screw using a 1/8” allen wrench (Figure. 3.2.3). This will allow access to
the seal nut. Using an adjustable wrench remove the seal nut by turning the nut
counter clock wise (Figure 3.2 4).

Figure 3.2.1

Figure 3.2.2

Flat Spot on
Liner