Error Indicators
Cytofuge
®
12 Cytocentrifuge error conditions are signaled by flashing codes appearing in the TIME
display on the front panel.
To clear an error condition, correct the problem and press the Stop/Open button.
Error Codes
Acceleration Error
Confirm no obstruction in bowl area, check
drive cone o-rings (see Section 5), contact
customer service.
Latch is not locked
Confirm cover is fully seated in closed
position.
Incorrect Speed
Remove and inspect rotor, check drive
cone o-rings (see Section 5), contact
customer service.
Cover is not closed
Confirm cover is fully seated in closed
position.
Out of balance
Confirm concentrators are balanced in
rotor and cytocentrifuge is on a level
surface.
Accessories
Product No.
Description
CC03
3-well Cell Concentrators and 3-well gaskets; Box of 10
CC01
1-well Cell Concentrators and 1-well gaskets; Box of 10
CLIP
Stainless Steel Clips for all Cell Concentrators; Bag of 10
GK01
Replacement Gasket for CC01 and CCR1
; Box of 100
GK03
Replacement Gasket for CC03 and CCR3
; Box of 100
CCR1
1-well Cell Concentrators, reusable, autoclavable; Box of 10
CCR3
3-well Cell Concentrators, reusable, autoclavable; Box of 10
BP01
Backing Plates for all Cell Concentrators; Box of 10
FF01
Filter Concentrators , Disposable; Box of 48
FF01-B
Filter Concentrators , Disposable; Box of 200 Filter Cards, 192 Concentrators
FFR1
Filter Concentrators, Reusable; Box of 200 Filter Cards, 20 Concentrators
FFCL
Stainless Steel Clips for Filter Concentrators; Box of 4
CFRT-12
Cytofuge 12 Rotor Assembly, replacement or spare, includes cover
CFRC-12
Cytofuge 12 Rotor Lid, replacement