Opening and Closing the Cover

An electronic operated cover interlock mechanism prevents operation if the cover is not completely
closed and locked, and prevents the cover from opening when the rotor is turning. An operating cycle can
be initiated only if the cover latch is engaged.

Cover Interlock By-pass

In case of power failure or malfunction, the Stop/Open button may not release the cover. In this event
the cover interlock mechanism can be released manually by inserting the manual latch release tool
supplied with the Cytofuge 12 into the small hole located in the front of the housing. Push the tool inward
to release the latch. Remove the tool before re-starting the cytocentrifuge.

CAUTION - The cover interlock by-pass is for emergency use only. Do not attempt to operate the
Cytofuge 12 while the latch release tool is inserted in the housing. Misuse of the interlock bypass
may expose the operator to unsafe conditions.

Using the Removable Rotor

The Cytofuge

®

12 Cytocentrifuge is shipped with the rotor installed and supported by a shipping

retainer. To prevent possible damage to the cytocentrifuge motor mount, it is important to remove the
rotor whenever the cytocentrifuge is moved or shipped to another location.

CAUTION - The rotor (CFRT-12) is intended for use with biological samples only. Do Not place
flammable or reactive substances in the rotor.

CAUTION - The rotor (CFRT-12) is designed for many years of service but, depending on the
frequency of use, the rotor components may develop cracks over time or if dropped on a hard
surface. Inspect the rotor for damage or cracks before each use. If damage is found discontinue
use and contact
customer service or your distributor for a replacement.

To remove the rotor:

Open cytocentrifuge cover. Grasp rotor lid from the black
center grip and lift straight up and out of cytocentrifuge.
Never lift the rotor by the center button.

To open the rotor:

With rotor removed from instrument, pull up on the center
button to release lid.

Pull up on the black center grip to remove the lid.

To re-install rotor:

Place lid onto rotor. Press center button to lock lid in place.

Lift rotor by black center grip. Place rotor assembly onto cone located in center of instrument.

CAUTION - Do not lift the rotor (CFRT-12) by the locking button - the lid may release allowing the
rotor bowl to fall. Always lift the rotor by the black center grip.