Iris Sample Processing StatSpin® CritSpin Microhematocrit System User Manual
Page 7

3
Error Indicators
Error codes for the CritSpin are specified by the Error/Service LED flashing or continuously
illuminated.
Error/Service =
Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of
18,000, which is the useful life of the drive system. Drive mechanism needs replacing. Contact an
authorized service center.
Error/Service light will flash when one of the following conditions occur:
1. System Error – Restart the centrifuge.
2. Motor Failure – Drive mechanism requires replacement.
3. Rotor Improperly Seated – Re-install rotor.
Error/Service and 120 s lights will flash when the cover is open during spin.
Accessories
Product No.
Description
RH12
Hematocrit rotor, 12-position
HR4C
Circular Hematocrit Reader
HR05
Card-style Hematocrit Reader
HT9H-10
Hematocrit Tubes, 40mm, glass, heparin
HT9U-10
Hematocrit Tubes, 40mm, glass, untreated
HP8U-10
Hematocrit Tubes, 40mm, plastic, heparin
HP8U-10
Hematocrit Tubes, 40mm, plastic, untreated
HS24-10
Clay Sealant Pad
CSD1
Digital Reader (included with CS22 Product Number)
A
B
C
120S
A Error/Service light
continuously
illuminated
B Error/Service
light flashing
120S
C Error/Service and
120 s flashing