Sensitivity Calibration

1. With the scale in regular weighing mode, press the Tare key to zero the

balance.

2. Press the Cal key.The balance will show “Scale”.

3. Press the Tare key. The balance will display one of the accepted

calibration weights.

Note: The balance can only be calibrated in grams.

4. Press the Mode key to toggle between available calibration weights.

5. Press Tare. The balance will flash “0” briefly and then flash the

calibration weight.

6. Place the correct amount of weight (in grams) on the platform and press

the Tare key.

7. The calibration weight will continue to flash briefly and then become

stable.

8. Once the weight is stable, remove the weight from the balance. The

scale has now been calibrated and is in regular weigh mode.

9. Press the Tare key if the scale does not return to zero.

Linear Calibration

1. In regular weighing mode, press the Tare key to zero the balance.

2. Press the Cal key. The balance will show “Scale”.

3. Press the Mode key. The balance will show “Line”

4. Press Tare. The balance will flash “0” briefly and then flash the first

calibration weight.

5. Place the correct amount of weight (in grams) on the platform and press

the Tare key.

6. The balance will flash the next calibration weight.

7. Place the appropriate amount of weight on the weighing platter and

press Tare.

8. The calibration weight will continue to flash briefly and then become

stable.

9. Once the weight is stable, remove the weight from the balance. The

scale has now been calibrated and is in regular weigh mode.

10.Press the Tare key if the scale does not return to zero.

Advertising