LPS Series User Instructions

5.

Press the HOLD or UNITS key to move through the list of choices. Press
TARE to accept the displayed value. . .

d. x is displayed. x = division size. Choices are 1, 2 , and 5. This means
the countby will be by 1, 2 or 5.

In the next parameter you will select the countby; for example:

1, 0.1, 0.01, etc.

Refer to the section

4.2

Determine the Display Division (n), Division Size (d) and

Decimal Point for information on choosing usable parameters.

6.

Press the UNITS key to move through the list of choices. Press TARE to
accept the displayed value. . .

dot. x is displayed. x = number of digits after the decimal point. Choices
are 0, 1, 2, or 3.

7.

Press the UNITS key to move through the list of choices. Press TARE to
accept the displayed value. . .

xxxxxx flashes on the display. This is the scale capacity calculated from
the choices you made for display divisions (n), division size (d) and
decimal point (dot).

8.

Perform step

8a

or

8b

.

8a.

Press the HOLD or UNITS key to return to step

3

and reset any of the

parameters in steps

4

-

7

.

OR

8b.

Press the TARE key to continue with calibration. . .

CAL.P0 is briefly displayed, showing the indicator is ready to acquire the
calibration zero point, then displays 0.0 weight with the last digit
flashing.

9.

Remove all weight from the scale and press the TARE key to acquire the zero
point.

The display will blink and then CAL.P1 is temporarily displayed, then
xxx is displayed (50% of capacity is the default value) with the last digit
blinking. This shows the scale is ready for the P1 calibration point.