5.10) P10.x: Calibration unit

x=0: kg as calibration unit

x=1: lb as calibration unit


As per the setting of P8, P9 and P10, following table is listed,
Table3: Kg calibration unit:

Display division value in different weight unit that can be used

Calibration
division value

kg lb Lb:oz

(oz)

0.0001kg 0.0001kg 0.0002lb Not

available

0.001kg 0.001kg 0.002lb Not

available

0.01kg 0.01kg 0.02lb 0.5oz

0.1kg 0.1kg 0.2lb 5

oz

1kg 1kg 2lb Not

available

10kg

10kg

20 lb

Not available

0.0002kg

0.0002kg

0.0005 lb

Not available

0.002kg

0.002kg

0.005 lb

0.1 oz

0.02kg

0.02kg

0.05 lb

1 oz

0.2kg

0.2kg

0.5 lb

Not available

2kg

2kg

5 lb

Not available

20kg

20kg

50 lb

Not available

0.0005kg

0.0005kg

0.001 lb

Not available

0.005kg

0.005kg

0.01 lb

0.2 oz

0.05kg 0.05kg 0.1

lb 2oz

0.5kg

0.5kg

1 lb

Not available

5kg

5kg

10 lb

Not available

50kg

50kg

Not available

Not available

Table4: Lb calibration unit:

Display division value in different weight unit that can be used

Calibration
division value

kg lb Lb:oz

(oz)

0.0001lb

Not available

0.0001lb

Not available

0.001 lb

0.0005 kg

0.001 lb

Not available

0.01 lb

0.005 kg

0.01 lb

0.2 oz

0.1 lb

0.05 kg

0.1 lb

2 oz

15

1 lb

0.5 kg

1 lb

Not available

10 lb

5 kg

10 lb

Not available

0.0002 lb

0.0001 kg

0.0002 lb

Not available

0.002 lb

0.001 kg

0.002 lb

Not available

0.02 lb

0.01 kg

0.02 lb

0.5 oz

0.2 lb

0.1 kg

0.2 lb

5 oz

2 lb

1 kg

2 lb

Not available

20 lb

10 kg

20 lb

Not available

0.0005 lb

0.0002 kg

0.0005 lb

Not available

0.005 lb

0.002 kg

0.005 lb

0.1 oz

0.05 lb

0.02 kg

0.05 lb

1 oz

0.5 lb

0.2 kg

0.5 lb

Not available

5 lb

2 kg

5 lb

Not available

50 lb

20 kg

50 lb

Not available

5.11) P11.x: select the weighing unit that may be chosen by pressing UNIT
x=0:

only

kg

x=1:

only

lb

x=2:

only

lb:

oz

x=3:

kg

or

lb

x=4:

kg

or

lb:

oz

x=5:

lb

or

lb:

oz

x=6: kg, lb, or lb: oz

Above setting of units may not be available as per above tables in 5.10


5.12) P12.x: Power-on zero-point range

x=0: the range is set to calibration zero point (CAL.P0) +1%FS

x=1: the range is set to calibration zero point (CAL.P0) +2%FS

x=2: the range is set to calibration zero point (CAL.P0) +5%FS

x=3: the range is set to calibration zero point (CAL.P0) +10%FS

x=4: the range is set to calibration zero point (CAL.P0) +20%FS

x=5: the range is set to calibration zero point (CAL.P0) +50%FS

x=6: the range is set to calibration zero point (CAL.P0) +100%FS