5.13) P13.x: Zero range for ZERO button after switch on

x=0: the range is set to calibration zero point (CAL.P0) +1%FS

x=1: the range is set to calibration zero point (CAL.P0) +2%FS

x=2: the range is set to calibration zero point (CAL.P0) +5%FS

x=3: the range is set to calibration zero point (CAL.P0) +10%FS

x=4: the range is set to calibration zero point (CAL.P0) +20%FS

x=5: the range is set to calibration zero point (CAL.P0) +50%FS

x=6: the range is set to calibration zero point (CAL.P0) +100%FS

x=7:

No

limitation

If zero key can be activated, it can clear the tare weight. If zero point is

above the setting range, the indicator will show “0¯ ¯ ¯ ¯”, and if zero point

is below the setting range, “0_ _ _ _” will be shown.


5.14) P14.x: select which zero point will be used when powered on and

weight signal is within the power-on zero-point range:

x=0: Choose current weight as current power-on zero point.

x=1: Choose calibration zero point as power-on zero point

x=2: Choose power-off zero-point as power-on zero point, and

power-off tare weight as current tare weight.


5.15) P15.x: select which zero point will be used when powered on and
weight

signal

is

NOT within the power-on zero-point range:

x=0: Choose current weight as current power-on zero point.

x=1: Choose calibration zero point as power-on zero point

x=3: Choose power-off zero-point as power-on zero point, and

power-off tare weight as current tare weight.

5.16) P16.x: Zero tracking range

x=0: 0d, means no tracking

x=1: +0.25d

x=2: +0.5d

x=3: +1d

x=4: +1.5d

x=5: +2d

x=6: +3d

x=7: +4d

x=8: +5d

Choose the zero tracking range as per the stability of weighing system,

accuracy and weight excursion. The normal setting is +0.5d ~ +1.5d.

17


5.17) P17.x: Data filter intensity

x=0 very weak

x=1

weak

x=2

middle

x=3

strong

The larger the chosen digit is, data filter intensity is stronger, and the speed of

data updating is lower. The normal setting is 2.

5.18) P18.x: Range of weight stability checking
x=0:

+0.5d

x=1…9:

+1d --- +9d

If the variety of weight reading is within the setting range continuously for

several times, the scale is recognized as stable. The normal setting is +1d ~

+3d.
5.19) P19.x: Overload range that can be displayed (when weight is lager than

range, “ˉˉˉˉˉ”will be shown):

x=0:

S+0d

x=1:

FS+9d

x=2:

101%FS

x=3:

102%FS

x=4:

105%FS

x=5:

110%FS

x=6:

120%FS

x=7:

150%FS

x=8:

200%FS

x=9:

No

limitation

========================================================



. The detail about RS232:

1. RS-232 connects between scale and Host:

Scale -----------------------Cable---------------------------Host

(DB9 female)--------(DB9 male)----(DB9 female)--------(DB9 male)

TXD 2 --------------- 2 ------------------------ 2 --------------2 RXD

RXD 3 --------------- 3 ------------------------ 3 --------------3 TXD

18