CONTENT

. SPECIFICATIONS. ……………………………………………… P1

. KEY FUNCTION. ……………………………………………… P3

. CALIBRATION. ………………………………….……………… P6

. Examination mode………………………………………………… P9

. SETUP MODE. …………………………………………………… P11

. Details of RS232 Communication………………………………... P18

. Calibration resolution, division and capacity …..………………… P21

. Meaning of some displayed symbols. ……………………………. P24

. Direction of indicator with bracket. ………………………. ……… P25

.Key Definition summary: ………………………. ………………...… P26

.Parameters setting summary: ………………………………... ...… P27

Operation Manual

Thank you for purchasing the indicator. Please read all operating instructions

carefully before use and keep the following points in mind:

* Avoid lengthy exposure to extreme heat or cold, your scale works best when operated
at normal room temperature. Always allow the scale to acclimate to a normal room
temperature before use
* Allow sufficient warm up time. Turn the scale on and wait for a few minutes if
possible, to give the internal components a chance to stabilize before weighing.
* These electronic scales are precision instruments. Do not operate near an in-use cell
phone, radio, computer or other electronic device. These devices emit RF and can
cause unstable scale readings. If your scale ever performs poorly, try moving the scale
to a different room or location.
* Avoid using in condition of heavy vibration and airflow.
* Read the weight reading in short time after loading. The output signature of load cell
and A/D may be little influenced after weighing for a long time.


.

SPECIFICATIONS:


·SCALE INDICATOR

1. Input signal range: 0mV

+30mV

2. Sensitivity: >0.2uV/grad
3. Internal Resolution: Approximately 520,000 counts
4. Display Resolution: can be selected between 500-100,000
5. System Linearity: within 0.01% of FS
6. Loadcell excitation Voltage: +4.4 VDC (MAX current: 55mA)
MAX 4- 350 ohm loadcells. “Only 140 indicator manual”
7. Calibration Method: Software calibration with long-term storage in EEPROM.


·SERIAL COMMUNICATIONS:

1. Mode: Full duplex or only output mode can be selected
2. Baud rate: 1200, 2400, 4800, 9600, or 19200 bps

3. Data format: 8NO, 7E1, 7O1

8data bits, non parity, 1 stop bit

7data bits, 1bit even or odd parity, 1 stop bit

4. Protocol: 7selected protocol (include the one compatible with NCI

1

2