0 model 3305 suggested test setup, Measuring electronic field shielding effectiveness – ETS-Lindgren 3303 Monopole Antennas User Manual

Page 21

Advertising
background image

5.0 Model 3305 Suggested Test Setup

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Measuring Electronic Field Shielding Effectiveness

Reduce amplifier power to zero watts before

band switching the Model 3303.

Reference:

• NSA 65-6

• MIL-STD-285

1.

Place the transmitting and receiving antennas so that they are

24 inches plus the thickness of the screen room wall apart from each

other.

2.

Set up equipment as shown in Suggested Test Setup Diagram on

page 22.

3.

Turn generator to ON position.

Disable RF output of generator.

Turn amplifier to ON position.

Turn receiver to ON position.

Turn Model 3301C power on.

Set receiver.

Set the attenuator at maximum (100 dB–150 dB) attenuation.

Enable RF output of the generator.

4.

Adjust signal amplitude on signal generator to maximum allowed by

amplifier input.

Model 3305 Suggested Test Setup

| 21

Advertising