0 installation and application, Connecting to power – ETS-Lindgren 3625-2 LISN User Manual

Page 13

Advertising
background image

www.ets-lindgren.com

Installation and Application

13

4.0 Installation and Application

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

The Model 3625/2 is provided with resistors to

help bleed off high voltage transients, but it is

advisable to connect the input and output

connectors to their proper power lines and

loads before connecting the monitor port to the

measurement instrumentation; otherwise,

power surges or transients can damage the test

instrumentation mixers or attenuators.

The Model 3625/2 Line Impedance Stabilization Network (LISN) is nominally

designed for a 25-ampere maximum load capacity, though this rating may be

increased to 30 amperes provided that the duty cycle is reduced to 20%.

Maximum line-to-line voltage must not exceed the 440 volts AC RMS.

Maximum line-to-ground voltage must not exceed 220 volts AC RMS.

Connecting to Power

The safety ground should be connected first

and disconnected last on the input side of the

unit.

The line side of the Model 3626/2 should be connected to the power source

using the three plug sockets provided. Attach these plugs to the pin receptacles

located on the back panel.

When the power is to be disconnected, remove the power source.