ETS-Lindgren 3850-2 LISN User Manual

Page 17

Advertising
background image

Installation

| 17

Spectrum Analyzer / Receiver

Use the BNC connectors on the front panel to connect the spectrum
analyzer or EMI receiver. See page 16 for a view of the front panel
connectors.

The monitor port which is not being monitored should be terminated
into a coaxial 50-ohm terminator.

Connect the input and output terminals to their proper mains
connections and EUT connections before connecting the monitor port
to the measurement instrumentation; otherwise, power surges may
damage the test instrumentation mixers or attenuators.

Removal of a terminator and connection to the BNC receptacle will not
generate power surges.

When the power is to be disconnected, remove the coaxial connection
to the measurement instrumentation first to avoid possible damage.