Operating Protocols

|

105

HI-60XX

S

ERIES

F

IELD

P

ROBES

Probe
Command

Description

Probe Response

B<CR>

Read battery voltage

:Bxx.xx<CR>

:B04.80

=100% of charge
=4.8 volts

:B04.00

=0% of charge
=4.0 volts

BP

Read battery capacity in
percentage

Bxx<N or F><CR>

:B64N

=100% of charge
=4.8 volts

:B00F

=0% of charge
=4.0 volts

N=safe operating level

F=fail level

Percentage reported in
hexadecimal, 0% to 64%

The

BP command is provided for backward compatibility and should

not be used to monitor the converter voltage, which always responds
with :B64N<CR>.