Hi-6113 laser data interface commands – ETS-Lindgren HI-6153 Electric Field Probe User Manual
Page 107

Operating Protocols
|
107
Probe
Command
Description
Probe Response
TF
Read temperature in
Fahrenheit
:Txxxx.<CR>
<
Null
><
CR
>
Send the ASCII null
character
:N<CR>
<Null><CR> is a special command that can be used as the initial
command to the probe
after it is turned on.
HI-6113
L
ASER
D
ATA
I
NTERFACE
C
OMMANDS
LDI
Command
Description
Probe Response
i
Laser data interface
identification string
:i6113<sr><sn><CR>
sr=10-character software
revision
sn=8-character serial
number
n
Read laser current
:nx.xxx.<CR>
o
Laser OFF command
:o
The laser and all LEDs except
the green Power LED will turn
off
r
Laser ON command
:r
The blue Laser LED will
illuminate immediately, then the
yellow Receive LED will
illuminate a few seconds after,
indicating the probe is ready for
operation