ETS-Lindgren 3301B Active Rod & Field Antenna (Archived) User Manual

Page 16

Advertising
background image

16

| Specifications

Saturation Indicator Impulse Response:

The saturation indicator will properly
indicate saturation for pulsed signals which
fall within the following boundaries:

1. The product of the pulse width to pulse

repetition rate must be greater than .003.


2. The pulse repetition rate must be less

than value listed for the applicable duty
cycle. See the following Table of
Acceptable Pulse Repetition Rates
.

5.1.1 Table of Acceptable Pulse Repetition Rates

Duty Cycle Maximum

PRF

10 19.7

MHz

20 14.0

MHz

30 9.2

MHz

40 5.6

MHz

50 420.0

kHz

60 130.0

kHz

70 87.0

kHz

80 57.0

kHz

90 30.0

kHz

5.2 Electrical Specifications

Input Impedance:

The input impedance determines the low
frequency roll of point. Depending on the
low frequency roll off selected, the input
impedance is >20 megohms, 1 megohm or
100 kilohm.

Output Impedance:

50 ohm (nominal)

Saturation Impedance:

A red LED on the front panel will illuminate
when the unit begins to saturate or clip. The
saturation indicator will indicate accurately
for any gain and attenuator setting.

Battery Low Indicator:

The power light on the front panel also
serves as a battery low Indicator. If the
green light does not come on when the
power switch is engaged, then the battery
requires recharging. It is recommended that
the Model 3301B be connected to the
battery charger at all times when not in use.

Archived 3/18/10