ETS-Lindgren 3301B Active Rod & Field Antenna (Archived) User Manual

Page 32

Advertising
background image

32

| Theory

of

Operation

high. However, the amplifier is being called upon to amplify signals at many

frequencies simultaneously. In the time domain, the amplifier would be seen to

be presented with quite a large power demand. The result is that impulsive

signals appear to saturate the amplifier at must lower signal levels. The

Model 3301B has been measured to begin going non-linear at 64 dBuV/m/MHz.

A saturation indicator is provided on the unit. This indicator will eliminate many

false readings due to non-linear operation. The saturation indicator has been

carefully tested to give an accurate, early warning of possible saturation.

However, some types of impulse signals may not trigger the indicator. The user

must, therefore, be careful to observe the published limits. Should these be

exceeded, the internal attenuation provided would relieve the problem.

Archived 3/18/10