Rice Lake Contego-Hygienic Weigh Module Kits, Lateral Cable Outlet User Manual
Page 19
Installation Manual
Hygienic Weighing Module Contego®
Sartorius
EN-17
Note:
The following operations must be performed at all supporting points (e.g. vessel foot) of
the weighing object (e.g. vessel)!
Procedure
Caution:
The following contact surfaces must be clean and even to ensure reliable sealing:
-
between upper plate and vessel foot
-
between lower plate, adaptor plate (if any) and foundation
• Check, if all threaded holes required for mounting (see Chapter 4.1.2) in the vessel foot (6) are
provided.
• Insert the O-ring (5) into the groove of the upper plate (4) and put the weighing module under the
vessel foot (6).
• Slip the bellow (not shown) over the mounting kit and the vessel foot (6) carefully so that the
mounting kit remains accessible for mounting.
• Fit the mounting kit to the vessel foot (6). It is essential to observe the property classes and tightening
torques of bolts (2) and washers (1); refer to Chapter 4.1.2.