Rice Lake Intercomp CS 750 - Users Manual User Manual
Page 14

CS750, Users
Rev R, May 2009
Page 14 of 27
Start up
1. Move the calibration blocking switch to the “CAL” position if you intend to
calibrate.
2. Turn scale power ON and wait for scale to warm up (3 minutes from power
on).
3. Press ZERO and hidden keys together and release to enter the calibration
mode.
First ten parameters
4. The scale shows “EE-EE
EE-EE
EE-EE
EE-EE”. Press the zero key. To skip to “EE-10
EE-10
EE-10
EE-10” enter a “1
11
1”.
To skip to “LL-00
LL-00
LL-00
LL-00” enter “2
22
2”. No skips will occur with an entry of “0
00
0”.
5. The scale shows “EE-00
EE-00
EE-00
EE-00”. Press the zero key. Enter the sample rate (1 to
64). The sample rate is the number of past readings that are averaged
together to make a reading.
6. The scale shows “EE-01
EE-01
EE-01
EE-01”. Press the zero key. Enter the update rate (1 to
32). The update rate is the speed at which the displayed weight is updated.
The smaller the number the faster the display will be updated.
7. The scale shows “EE-02
EE-02
EE-02
EE-02”. Press the zero key. Enter demand versus
continuous on the optional serial output. For this product, you should always
set this to "Continuous" (0).
Setting
Type
0
Continuous on
1
Continuous off
8. The scale shows “EE-03
EE-03
EE-03
EE-03”. Press the zero key. Enter the baud rate of the
serial output (0 to 7).
Setting
Baud Rate
0
9600
1
4800
2
2400
3
1200
4
600
5
300
6
150
7
75
8
19.2K
9
38.4K