Performance test successfully completed – Rice Lake MA37 Sartorius Moisture Analyzer User Manual
Page 41

41
MA Performance Test
Operating Instructions Moisture Analyzer MA37
MA performance test
Device is tested with reference sample.
Temperature: 70 °C
Duration: 05:00 min
00:13 min
y The display shows the progress of the test and the remaining time .
Performance Test Measurement Not Successfully
Completed
t
MA performance test
Measurement canceled
y The message “
Measurement canceled
” appears .
t
Select
l
to confirm .
y
The display switches back to the menu .
If a fault has been detected:
t
MA performance test
Device is not OK.
5 MApi << 10 MApi
Temperature 70 °C
Please contact Sartorius service
a
y The message “
Device is not OK
” appears .
t
Select
m
to confirm .
Contact Sartorius service if the message “
Device is not OK
” appears .
Performance Test Successfully Completed
a
l
MA performance test
Device is OK.
Result 10 MApi
Temperature 70°C
3
y If the test is successful, the message “Device is OK” first appears . Then the
analyzer indicates that the sample can be removed .
Remove old sample
t
Select
l
to confirm .
y You can remove the sample and return to the menu .