8-2. Initializing the Balance
This function returns the following parameters to factory settings.
Calibration
data
Function
table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External calibration weight and target weight value
Function switch settings
Note Be sure to calibrate the balance after initialization.
Step 1 Press the
ON:OFF
key to turn off the display.
Step 2 While pressing and holding the
key and
the
SAMPLE
key, press the
ON:OFF
key to
display
p5
.
Step 3 Press the
SAMPLE
key to display
Clr
.
Step 4 Press the
key.
To cancel this operation, press the
CAL
key.
Step 5 Press the
RE-ZERO
key to display
Clr go
.
Step 6 Press the
key to initialize the balance.
The balance will automatically return to the weighing
mode.
en d
0 .0 0 0
g
Cl r n o
Cl r
p5
Cl r g o
With these keys Press the key
held down