Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual
Page 8

6
2-1. Auto-centering Pan (AX-MC10K/30KPAN) Installation Procedure
Caution: During installation, do not to apply excessive force to the balance.
When used as a mass comparator, the balance can achieve even more precise
weighing, by using the auto-centering pan (sold separately), which reduces
eccentric loading errors.
1.
Place the pan support on the balance. Place the breeze break provided with the
AX-MC10K/30KPAN option on the balance.
Place the auto-centering pan guide on the pan support, and then place the auto-centering
pan on the auto-centering pan guide.
Note
When the weighing pan is replaced
with the auto-centering pan, be
sure to calibrate before weighing.
Refer to “7. CALIBRATION”.
Auto-centering pan
Auto-centering pan guide
Option AX-MC10K/30KPAN breeze break
Pan support