Appendix E: Configurations and test protocols
Note
The protocols Included are DP 4s (default) and DP 2s.
Name
No. of
Frequenci
es
F2 Frequency
L1/L2
Maximum Average
Time
Pass
SNR
# Passing
Frequency for
Test Pass
DP 4s
4
2.0, 3.0, 4.0, 5.0 kHz 65/55
4 seconds
6 dB
3
DP 2s
4
65/55
2 seconds
6 dB
3
Note
The DPOAE protocols differ in the amount of averaging time per frequency.
The DP 4s protocol is the default. This averaging time is the time the
screener uses to render a decision on that frequency. The DP 2s (2 second)
protocol uses 2 seconds per frequency and results in faster test times at a
minimum of 8 seconds. However, it may increase the false referral rate.
The DP 4s (4 second) protocol uses 4 seconds averaging time per
frequency and results in slower test time at a minimum of 16 seconds per
tes. However, it may reduce the false referral rate.
56 Appendices