Chapter 3 Patient monitoring
45
7.
Eject the probe cover by pressing the ejection button, and hygienically dispose of it.
8.
Return the probe to the probe well.
Monitored temperature measurement
The procedures for monitored and predictive temperature measurements are the same,
with the following exceptions:
For monitored measurements:
•
The monitor must be set to take a monitored temperature. (See
temperature type and measurement units”
•
The monitor displays the temperature continuously.
•
The measurement continues until the probe is replaced in the probe holder.
WARNING If the probe becomes contaminated, follow the instructions under
“Thermometer and probe cleaning procedure”
Caution Do not monitor temperature continuously for more than 5 minutes.
Note
Verify that the temperature measurement type is set to monitored. (The letter
, to the right of the temperature display window and below
or
,
is illuminated.)