Nurse call – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 55

Directions for Use
Chapter 4 Alarms and alerts
51
Nurse call
The monitor can be connected to a Nurse Call system through a customized cable that
connects to the Nurse Call connector. When the cable is connected and operational, the
monitor immediately notifies the Nurse Call system when a patient alarm occurs.
Alarm limits for vital signs
Vital sign
Low limit (default)
High limit (default)
Resolution
Systolic (mmHg)
Neonatal
30 - 110 (50)
35 - 115 (100)
5 mmHg (0.5 kPa)
Pediatric
35 - 150 (75)
40 - 155 (145)
5 mmHg (0.5 kPa)
Adult
35 - 250 (75)
40 - 255 (220)
5 mmHg (0.5 kPa)
Diastolic (mmHg)
Neonatal
15 - 95 (30)
20 - 100 (70)
5 mmHg (0.5 kPa)
Pediatric
20 - 120 (35)
25 - 125 (100)
5 mmHg (0.5 kPa)
Adult
25 - 225 (35)
30 - 230 (110)
5 mmHg (0.5 kPa)
Mean arterial pressure (mmHg)
Neonatal
15 - 100 (35)
20 - 105 (80)
5 mmHg (0.5 kPa)
Pediatric
20 - 130 (50)
25 - 135 (110)
5 mmHg (0.5 kPa)
Adult
25 - 245 (50)
30 - 250 (120)
5 mmHg (0.5 kPa)
Pulse rate (beats per minute)
Neonatal
35 - 210 (100)
40 - 215 (200)
5 bpm
Pediatric
35- 210 (50)
40 - 215 (150)
5 bpm
Adult
35 - 210 (50)
40 - 215 (120)
5 bpm
SpO
2
(%)
Neonatal
50 - 98 (85)
51 - 99 (95)
1%
Pediatric
50 - 98 (85)
51 - 99 (---)
1%
Adult
50 - 98 (85)
51 - 99 (---)
1%
Nurse Call Cable
Connector