Low battery alerts – Welch Allyn AcuityLink Clinician Notifier - User Manual User Manual
Page 51

Directions for Use
Respond to alarms and alerts
47
Low battery alerts
Mobile device low battery alerts are associated with a single beep tone, followed by a
long pause.
Main Battery Low alerts
To Respond to a Main Battery Low popup window
1.
In the popup window, tap OK.
2.
If you do not see a Clinician Notifier program window on your mobile device screen,
tap
, and then tap
Clinician Notifier.
A Clinician Notifier screen appears.
3.
Tap
in the upper right corner of the screen.
A confirmation popup window appears.
4.
Tap Yes.
You are logged out of the Clinician Notifier program.
WARNING Connecting mobile devices to the Internet, or running too many
programs in the foreground or background while the Clinician Notifier program is
running, can greatly reduce the battery life and memory of the devices, thereby
compromising alarm delivery.
WARNING Always recharge a mobile device when it is not in use and at first
notification of low batteries. A device with low batteries might not indicate or
appropriately clear all of a patient’s alarms. Allowing batteries to fully discharge
can result in depletion of the backup battery and loss of Clinician Notifier
configurations.