Welch Allyn CP 150 12-lead resting electrocardiograph v.1.00.XX - User Manual User Manual

Page 54

Advertising
background image

Item

Specification

6x2-5.0s @ 25 mm/s
6x2-5.0s @ 50 mm/s
12x1-10.0s @ 25 mm/s

Report formats, internal printer,

Average

3x4+3R @ 25 mm/s
3x4+3R @ 50 mm/s
6x2+1R @ 25 mm/s
6x2+1R @ 50 mm/s
No print

ECG storage (in test directory)

At least 100 ECG tests

Patient storage

Up to 50 patients

Frequency range

0.3 to 150 Hz

Digital sampling rate

> 1,000 samples/second/channel

Pacemaker detection

ANSI/AAMI EC11

Power requirement

Universal AC power supply 110-240 V~, 50/60 Hz ~, 1.5 A maximum

AC fuses

Time-lag type, 2.0-amp 250-V rating, Littelfuse 0218002P or equivalent

Rechargeable battery

9 cells

Rating

10.8 V 6.75Ah (73Wh)

Composition

Lithium-ion

Recharge time to 100 percent

capacity

4 hrs

Full charge capacity

25 ECG tests @ 20 minutes per test

8 hrs of continuous operation or 250 continuous ECGs

Filters

High-performance baseline

0.5 Hz

Muscle tremor

35 Hz

AC interference

50 Hz or 60 Hz

Safety, EMC, and regulatory

compliance

ANSI/AAMI EC11**
CAN/CSA C22.2 No. 601.1
CAN/CSA C22.2 No. 601.1.2

IEC/EN 60601-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4

50 Specifications

CP 150 12-lead resting electrocardiograph