Welch Allyn CP 150 12-lead resting electrocardiograph v.1.00.XX - User Manual User Manual

Page 55

Advertising
background image

Item

Specification

CAN/CSA C22.2 No. 601.1.4
CAN/CSA C22.2 No. 601.2.25

IEC/EN 60601-1-6
IEC/EN 60601-2-25 ***
IEC/EN 60601-2-51* (3x4 report format)

Standard connectivity

1 USB client
4 USB hosts

Connectivity with electronic

medical records

Through the Welch Allyn CardioPerfect workstation software

Electrodes

Rigorously tested for conductivity, adhesion, and hypoallergenic qualities, and exceed

all AAMI standards

Power cable

Meets or exceeds Type SJT.

Environmental operating

conditions

Temperature

+10° C to +40° C (+50° F to +104° F)

Relative humidity

15 - 95% noncondensing (30 - 70% for printing)

Atmospheric air-pressure limits 700 - 1060 hPa

Environmental storage

conditions

Temperature

-20° C to +50° C (-4° F to +122° F)

Relative humidity

15 - 95% noncondensing

Atmospheric air-pressure limits 700 - 1060 hPa

Protection against electric shock Class I, internally powered Type CF

Mode of operation

Continuous

*

If you print at a high gain setting, the waveform or calibration marks might be clipped. This clipping does not

comply with clause 51.103.1 of the IEC/EN 60601-2-51 standard. Use a lower gain setting to observe the full

waveform.

**

Per AAMI EC11:1991/®)2007 Diagnostic Electrocardiographic Devices, Section 3.1.2.1 Disclosure of cautionary

information/ performance characteristics paragraph c) Accuracy of input signal reproduction, the manufacturer

shall disclose the methods used to establish overall system error and frequency response. Welch Allyn has used

methods A & D, as prescribed in section 3.2.7.2 and 4.2.7.2 of this same standard, to verify overall system error

and frequency response. Because of the sampling characteristics and the asynchronism between sample rate and

signal rate, digital ECG systems such as the CP 150 may produce a noticeable modulating effect from one cycle to

the next, particularly in pediatric recordings. This phenomenon is not physiologic.

Directions for use

Specifications 51