P= 30w 60 rpm esc disp start – Welch Allyn Corival V2 - User Manual User Manual
Page 24

Operator Manual
Corival
24
Whenever the test subject is allowed to or has to stop the exercise test, the recovery button has
to be pressed. The following will appear on screen:
In this example the recovery phase will bring the workload down to 50 watt. The timer is
showing 10:30 seconds and will go on with the registration during the recovery as long as the
RPM is above 30 (the BPM can still be measured when BPM is displayed at the confirmation
button). To stop the recovery phase press the END button to go back to the PROTOCOL
menu. In the menus MANUAL, TERMINAL or ANALOG the BP function is the same as
described above.
B.3.7.5
Protocol delete
This function allows you to delete a protocol. Select the protocol which you do want to
remove (in this example TESTPROTOCOL) and confirm with OK. The following screen will
appear:
Confirm with OK when you agree and the message [deleted] will appear on the screen
Rec P= 50W t=10:30
P= 50W 60 rpm
θ = 174/min
esc esc
testprotocol
Delete?
esc ok
BPM
p = 123 mmHg
α Ԁ
P= 30W 60 rpm
esc DISP START