Recommended separation distances between portable and mobile RF

communications equipment and the Welch Allyn CardioPerfect Pro

Cardiograph

The Welch Allyn CardioPerfect Pro Cardiograph is intended for use in an electromagnetic environment in which

radiated RF disturbances are controlled. The customer or the user of the Welch Allyn CardioPerfect Pro Cardiograph

can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF

communications equipment (transmitters) and the Welch Allyn CardioPerfect Pro Cardiograph as recommended

below, according to the maximum output power of the communications equipment.

Rated maximum

output power of

transmitter

W

Separation distance according to frequency of transmitter

m

150 KHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

0,01

0.12

0.12

0.23

0,1

0.37

0.37

0.74

1

1.2

1.2

2.3

10

3.7

3.7

7.4

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.