Figure 7.1 The Pre-calibration dialog box


Fill in the appropriate settings. For a description of the options see the following table.

Setting

Description

Lot code

Enter the Lot code for the Flow transducers located on the box the
Transducers came in.

Calibration
code

Enter the Calibration code for the Flow transducers located on the box the
Transducers came in. See example of a schematic cutout of the label from
the Flow transducers box below.

Note: the sensor calibration information can also be set in the
Spirometry settings (recording tab). Please make sure that Lot code and
Calibration code are still accurate before calibrating.

Vica test:
If the VCT400 was selected the Lot Code and Calibration Code options
are not applicable and grayed out.

Syringe Volume Select the appropriate Syringe Volume.


Tip: See page 20 on how to change the default setting.

Current
Calibration
Factor

This value cannot be changed and indicates the correction factor applied
to the calibration data from the previous session. Once the calibration is
performed, this value will be updated. The factor displayed is the average
of the inspiration and expiration calibration factor.