Cardioperfect workstation, Spiroperfect module – user manual – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual

Page 65

Advertising
background image

CardioPerfect Workstation

SpiroPerfect Module

– User Manual

DIR 80012334 Ver. F

65 / 95


Report Area

Description

Date and time of Pre and Post efforts.

Norm Profile used to calculate
predicted values.

For each selected Pre and Post effort,
this section provides quality indication
regarding individual effort
acceptability.

Variance of best FVC and FEV1 in Pre
and Post phases. ATS standards
require this variance to be <150 mL.
Indication of Pre and Post effort
repeatability being met.
If criteria is not met, the reason why is
provided.

Main results of the Spirometry test.
See section 5 for settings that
influence the amount of information
shown here.

Spirometry interpretation area.

Tabular presentation of a subset of
measurement data.


Advertising