Chapter 5 Performing ECG Tests

Welch Allyn CP 100 Electrocardiograph

Figure 33. Auto ECG Acquisition” Screen

4.

If prompted, choose whether to wait for the electrocardiograph to acquire 10 seconds
of filtered, processed data before printing.

If you override the wait time and print the available data immediately, be aware that
the printed data will be insufficient in quality or quantity or both.

The report prints.

After printing, the Auto ECG Post-Print screen appears.

Figure 34

on page 51.

Note

If you do not enter the patient data, write the patient identification information on
the printed report immediately after the ECG test to avoid associating the ECG
report with the wrong patient.

ECG report Print Copy feature will not be available if no Patient Name or ID are
entered.

If a red stripe appears along the edge of your report, replace the paper. See

“Loading the Thermal Chart Paper”

on page 21.

Doe, Jane

9:17AM Oct 16 05

Auto ECG Acquisition

Collecting Data

The electrocardiograph displays messages here
when it is doing the following:

Collecting data

Analyzing data

Printing the report

Doe, Jane

9:17AM Oct 16 05

Auto ECG : Acquisition

Printing Report

Gain: 10 mm/mV
Speed: 25 mm/s
Baseline Filter: ON
Muscle Filter: ON

HR 120